Validação de método para determinação de ácidos orgânicos voláteis em efluentes de reatores anaeróbios empregando cromatografia líquida
AUTOR(ES)
Cerqueira, Maristela Barnes Rodrigues, Dias, Adriana Neves, Caldas, Sergiane Souza, Santana, Fabrício Butierres, D'Oca, Marcelo Gonçalves Montes, Primel, Ednei Gilberto
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2011
RESUMO
This work deals with the method validation for the determination of acetic, propionic and butyric acids (VFAs) in wastewaters from anaerobic reactors by HPLC-DAD. Separation was performed using a C18 column and the mobile phase composition were water pH 3.0 and methanol 90:10 (v/v). The detection and quantification was carried out at 220 nm. The method shows good linearity (r²>0.996), with adequate accuracy (89-102%) and relative standard deviations lower than 18%. The matrix effect was considered low (-4.1, -3.9 and 1.4%). The developed method is fast, simple and cheap; and it was applied in wastewater samples from anaerobic reactor.