Latex enzyme immunoassay for measuring IgG antibodies to rubella virus.
AUTOR(ES)
Duverlie, G
RESUMO
A latex enzyme immunoassay (LEIA) for detecting IgG class antibody to rubella virus was compared with the latex agglutination test (LA) and a haemagglutination inhibition assay (HI). Of 243 sera tested, four discrepant results were observed among all three techniques, corresponding to borderline values. Except for one sample containing specific IgM class antibody, the difference between quantitative results from each pair of tests was always within a value corresponding to two doubling dilutions and was considered to have acceptable variation. The LEIA technique allowed 10 seroconversions to be detected, as determined by the other techniques. The LEIA required a single 1 in 8 sample dilution, took 30 minutes, and provided a useful alternative assay for the quantitation of rubella antibodies.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=502758Documentos Relacionados
- Enzyme immunoassay for detection of immunoglobulin M (IgM) and IgG antibodies to Mycoplasma pneumoniae.
- Monoclonal antibody-based capture enzyme immunoassays for specific serum immunoglobulin G (IgG), IgA, and IgM antibodies to respiratory syncytial virus.
- Evaluation of Cobas Core Rubella IgG EIA recomb, a new enzyme immunoassay based on recombinant rubella-like particles.
- Preliminary multiplex microarray IgG immunoassay for the diagnosis of toxoplasmosis and rubella
- Automated microparticle enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii.