2021-06

Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins

ABSTRACT Objectives The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in samples with the MCHC > 370 g/L, in the presence of cold agglutinins. Methods In this study, 60 blood samples (group 1) with the MCHC < 370 g/L (without cold agglutinins) and 78 blood samples (group 2) with the MCHC > 370 g/L (with cold agglutinins) were used to compare the two analytical channel...