2015-02

The Electrochemical Behavior of the NiTi Alloy in Different Simulated Body Fluids

In order to improve the NiTi alloy biocompatibility, surface treatments become very important. Nevertheless, researchers use different solutions to simulate the body fluids in electrochemical assays, and the correlation between the obtained results is difficult and might not even be possible. The present paper evaluated the electrochemical behavior of polished NiTi surfaces exposed to different simulated body fluid solutions: Hanks solution, Hanks’ balanced salt (HBSS) solution, saline body fluid (SBF) solution, and Ringer solution. The electrochemical behavior of NiTi was evaluated by open ...