2010

Photometric procedure for quantitative analysis of Aflatoxin B1 in peanuts by thin-layer chromatography using charge coupled device detector

A photometric procedure was developed for determination of aflatoxin B1 in peanuts by TLC-CCD technique. The quantification and detection limit were 1.2 μg kg-1 and 0.4 ng per spot, respectively, with mean recovery of 98%. The CCD camera is sufficiently sensitive to detect small changes in spots fluorescence intensity and the results for performance confirmed the efficiency of the method. Another important property of CCD detector is its linearity for a wide range of luminous stimulus determined by analysis of five-point calibration curves using the intensity of AFB1 fluorescence versus AFB1 ...

Texto completo