02/09/2019

Iatrogenic foreign body in urinary bladder: Holmium laser vs. Ceramic, and the winner is…

ABSTRACT Introduction Urological surgery is estimated to be the third most common cause of iatrogenic-retained foreign bodies 1. Presentation A 76-year old man was undergoing a transurethral resection of bladder tumor with a 26-Ch continuous flow resectoscope (Karl Storz, Germany). Before starting resection, a detachment of resectoscope sheath tip was noted. The ceramic tip was free-floating in the bladder lumen, and it would not fit within the sheath, making direct extraction using the loop impossible. An attempt was made to break it with a stone punch, but it was unsuccessful due to impos...