HPLC METHOD IMPROVEMENT FOR 4-NONYLPHENOL DETERMINATION IN ANAEROBIC REACTORS: DECREASING SOLVENT CONSUMPTION AND WASTE GENERATION

The aim of this study was to develop a chromatographic method for 4-NP determination in anaerobic batch reactors using low amount of organic solvents and also evaluate 4-NP degradation in the presence of different co-substrates. Chromatographic parameters was improved for stationary phase (C-8 column), mobile phase (acetonitrile/water solution 90:10, v:v), column temperature (40 ºC), eluent flow (0.8 mL min-1), injection volume (100 µL) and wavelengths in fluorescence detector (FLD) (λex=225 nm and λem=305 nm). Solid phase extraction (SPE) was performed for sample purification with smaller...