2009

Development and validation of a GC-FID assay for determination of fluvastatin in pharmaceutical preparations

A gas chromatographic method has been developed for the assay of fluvastatin sodium (FLU). FLU was silylated with N,O-bis(trimethylsilyl)trifluoroacetamide-1% trimethylchlorosilane at 90 ºC for 30 min and analysed in a DB-1 column by capillary gas chromatograph with a flame ionization detector. The method was validated. The assay was linear over the concentration range at 10.0 to 50.0 µg mL-1. The limit of detection and the limit of quantitation were 1.0 and 3.0 µg mL-1, respectively. The recoveries of FLU derivatives were in the range of 99.25-99.80%. In inter-day and intra-day analysis, t...