Determinação de bifenilos policlorados em soro de cordão umbilical através de extração por hidrólise ácida seguida de cromatografia a gás acoplada a um microdetector de captura de elétrons

The present work describes the determination of polychlorinated biphenyls in 123 umbilical cord serum samples by liquid-liquid extraction method with acid hydrolyze step and analysis by GC-mECD. The analytical method was evaluated with following figures of merit for all PCBs: linearity (>0.997); precision (<12.55%); recoveries (73-119%); limit of detection (0.1 ng mL-1); limit of quantification (0.25-0.5 ng mL-1). The results showed high contamination in the analyzed samples. PCB more frequent was 138 (66.7%), followed by 180 (55.3%) and 52 (51.3%).