A sensitive bioanalytical method development and validation of cabozantinib in human plasma by LC-ESI-MS/MS

ABSTRACT A simple, sensitive and specific liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the quantification of cabozantinib (CZ) in human plasma using cabozantinib-d4 (CZD4) as an internal standard (IS). Chromatographic separation was performed on Xbridge C18, 50 x 4.6 mm, 5 mm column with an isocratic mobile phase composed of 10mM Ammonium formate and Methanol in the ratio of (20:80 v/v), at a flow-rate of 0.7 mL/min. CZ and CZD4 were detected with proton adducts at m/z 502.2 ® 391.1 and 506.3 ® 391.2 in multiple reaction monitoring (MRM) positive mode re...