Validação de método analítico livre de acetonitrila para análise de microcistinas por cromatografia líquida de alta eficiência
AUTOR(ES)
Cassini, Sérvio Túlio Alves, Antunes, Paulo Wagner Pereira, Keller, Regina
FONTE
Quím. Nova
DATA DE PUBLICAÇÃO
2013
RESUMO
Blooms of cyanobacteria represent a public health risk due to their cyanotoxins such as microcystins. Liquid chromatography techniques to separate and quantify microcystins invariably use acetonitrile as the organic component of the mobile phase. The price and availability of acetonitrile together with its elevated toxicity encourage the validation of acetonitrile-free methods of microcystin analysis. In this work, methanol was employed as the organic solvent of the mobile phase and the validation method was performed with different environmental water samples. The method showed limits of detection between 0.17 and 0.25 µg/L and of quantification between 0.55 and 0.82 µg/L for the microcystin variants: -RR, -YR, -LR, -LA.
Documentos Relacionados
- Desenvolvimento e validação de método analítico para nistatina creme vaginal por cromatografia líquida de alta eficiência
- Validação de método analítico por cromatografia líquida de alta eficiência para doseamento do adapaleno em suspensões de nanocápsulas
- Determinação do ácido trans,trans-mucônico em urina: validação de um método analítico por cromatografia líquida de alta eficiência
- Recuperação e reutilização de acetonitrila utilizada em análise de açúcares por cromatografia líquida de alta eficiência.
- Método analítico para a determinação de meloxicam em plasma humano por cromatografia líquida de alta eficiência (CLAE)