Sistema de análises em fluxo polivalente para a determinação espectrofotométrica de fármacos
AUTOR(ES)
Schmidt Jr, Erlando, Melchert, Wanessa R, Rocha, Fábio R. P
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2011
RESUMO
A flow system based on the sandwich technique is proposed for the sequential determination of ascorbic acid, dipyrone, acetylcysteine, captopril and paracetamol. The procedure is based on the reduction of Cu(II) by the analytes followed by the spectrophotometric measurement of the complex of Cu(I) with 2,2'-biquinoline 4,4'-dicarboxylic acid. Linear responses were achieved in the µmol L-1 range, with coefficients of variation better than 1.7%. Sampling rate was estimated as 60 determinations per hour, consuming 230 µg of BQA and generating 2.5 mL of waste per determination. Results for commercial samples agreed with those obtained by procedures recommended by the American and European pharmacopeias at the 95% confidence level.