Real-Time PCR for Quantification of Human Herpesvirus 6 DNA from Lymph Nodes and Saliva

AUTOR(ES)
FONTE

American Society for Microbiology

RESUMO

A real-time quantitative PCR assay has been developed to measure human herpesvirus 6 (HHV-6) DNA in biological specimens. The assay sensitivity was 10 copies of DNA per well, with a linear dynamic range of 10 to 107 copies of HHV-6 DNA. Intra- and interassay variations were, respectively, 0.88 and 0.8% for samples containing 102 DNA copies, 0.99 and 0.96% for samples containing 104 copies, and 0.76 and 0.9% for samples containing 106 copies. Among 34 saliva samples from healthy subjects, 26 were found to contain HHV-6 DNA (76.5%; median, 23,870 copies/ml), and following a single freeze-thaw cycle, 25 of the same samples were found to be positive for HHV-6 DNA, although at a statistically significantly lower concentration (median, 3,497 copies/ml). The assay enabled detection of HHV-6 DNA in lymph node biopsies from patients with Hodgkin's disease (HD) (13 of 37 patients [35.1%]), B-cell neoplasms (8 of 36 patients [22.2%]), and T- or NK-cell neoplasms (3 of 13 patients [23.1%]), with concentrations ranging from 100 to 864,640 HHV-6 copies per μg of DNA (HHV-6B being found in every case except two). All HD patients infected with HHV-6 presented clinically with the nodular sclerosis subtype of HD. The real-time quantitative PCR assay developed here was simple to perform and was sensitive over a wide range of HHV-6 concentrations. It therefore appears to be of potential value in clinical investigation or diagnosis of HHV-6 infection.