Rapid automated identification of gram-negative bacilli from blood cultures with the AutoMicrobic system.
AUTOR(ES)
Schifman, R B
RESUMO
Automated identification of gram-negative bacilli directly from blood culture bottles by using the AutoMicrobic System (AMS) was evaluated with a modified procedure for the AMS Enterobacteriaceae-plus nonfermenter identification card. A total of 150 strains were tested (44 clinical and 106 seeded) and compared with a conventional identification procedure. These strains included 107 Enterobacteriaceae and 43 oxidase-positive or glucose-nonfermenting, or both, organisms. AMS identifications on one of these strains were not interpretable owing to equal probability AMS identification values. Of the remaining 149 strains, 138 (92.6%) were correctly identified within 8 to 13 h of the first reading. Of 69 identifications analyzed after 6 h of incubation, 91% were correct. This procedure was found to be rapid, convenient, and nonlabor intensive and is recommended for presumptive identification of gram-negative bacilli in blood cultures.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=272072Documentos Relacionados
- Rapid identification and antimicrobial susceptibility testing of gram-negative bacilli from blood cultures by the AutoMicrobic system.
- Rapid antimicrobial susceptibility testing of Gram-negative clinical isolates with the AutoMicrobic system.
- Evaluation of the AutoMicrobic system for identification and susceptibility testing of gram-negative bacilli.
- Direct identification and susceptibility testing by the AutoMicrobic system of gram-negative bacilli from urine specimens.
- Comparison of the autoSCAN-W/A rapid bacterial identification system and the Vitek AutoMicrobic system for identification of gram-negative bacilli.