Quantitation of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Cell-Free Cervicovaginal Secretions: Comparison of Reverse Transcription-PCR Amplification (AMPLICOR HIV-1 MONITOR 1.5) with Enhanced-Sensitivity Branched-DNA Assay (Quantiplex 3.0)

AUTOR(ES)
FONTE

American Society for Microbiology

RESUMO

Two commercially available hypersensitive assays for human immunodeficiency virus type 1 (HIV-1) RNA quantitation, AMPLICOR HIV-1 Monitor Test 1.5 and Quantiplex HIV RNA 3.0, were compared to detect and quantify HIV-1 RNA in the cell-free fraction of cervicovaginal secretions collected by vaginal washing. Three panel specimens were used: pooled cervicovaginal secretions spiked with HIV-1 subtype A or HIV-1 subtype B and cervicovaginal lavages from HIV-positive and HIV-negative women. Compared to the AMPLICOR HIV-1 Monitor Test 1.5 assay, the Quantiplex HIV-1 3.0 assay yielded higher estimates of HIV-1 RNA concentrations in several tested samples spiked with HIV-1 RNA subtype A, as well as subtype B, particularly samples containing low amounts of HIV-1 RNA. The sensitivity and specificity of the AMPLICOR HIV-1 Monitor Test 1.5 assay were 93 and 100%, respectively; the sensitivity and specificity of the Quantiplex HIV RNA 3.0 assay were 97 and 50%, respectively. In conclusion, in quantifying HIV-1 RNA in cervicovaginal secretions, the Quantiplex HIV RNA 3.0 may lack specificity, and the AMPLICOR HIV-1 Monitor Test 1.5 assay, although highly specific, may lack sensitivity.

Documentos Relacionados