Preload and torque removal evaluation of three different abutment screws for single standing implant restorations
AUTOR(ES)
Stüker, Rafael Augusto, Teixeira, Eduardo Rolim, Beck, João Carlos Pinheiro, Costa, Nilza Pereira da
FONTE
Journal of Applied Oral Science
DATA DE PUBLICAÇÃO
2008-02
RESUMO
Several authors still consider the mechanical problems of fracture and component loosening as the main causes of failure of implant-supported restorations. The purpose of this in vitro study was to compare the preload of three types of screw for transmucosal abutment attachment used in single implant-supported prosthesis through strain gauge and removal torque measurements. Three external hex fixtures were used, and each received a transmucosal abutment (Cera One®), which was fixed to the implant with its respective screw: Group A- gold screw, Group B- titanium screw and Group C- surface-treated titanium screw (Ti-Tite®). Ten screws of each type were attached applying a 30.07±0.28 Ncm torque force and maintained in position for 5 minutes. After this, the preload values were measured using strain gauges and a measurement cell. Gold screws presented higher preload values (131.72±8.98 N), followed by surface-treated titanium screws (97.78±4.68 N) and titanium screws (37.03±5.69 N). ANOVA (p<0.05) and Tukey's test (p<0.05) were applied. Statistically significant differences were found among the groups for both preload and removal torque values. In conclusion, gold screws may be indicated to achieve superior longevity of the abutment-implant connection and, consequently, prosthetic restoration due to greater preload values yielded.
Documentos Relacionados
- Relation between implant/abutment vertical misfit and torque loss of abutment screws
- Removal torque of zirconia abutment screws under dry and wet conditions
- Preload loss and bacterial penetration on different implant-abutment connection systems
- Effects of different torque levels on the implant-abutment interface in a conical internal connection
- Evaluation of screw loosening on new abutment screws and after successive tightening