Physiological importance of different atrioventricular intervals to improved exercise performance in patients with dual chamber pacemakers.
AUTOR(ES)
Haskell, R J
RESUMO
To determine the importance of different atrioventricular intervals during exercise in patients with dual chamber pacemakers, seven patients with complete heart block and sinus rhythm were exercised in different pacing modes and atrioventricular intervals: (a) ventricular inhibited (VVI) pacing with no synchronous atrial augmentation or rate responsiveness; (b) atrial synchronous ventricular or DDD pacing with a short mean (SD) atrioventricular interval of 66 (4) ms; and (c) DDD pacing with a long atrioventricular interval of 168 (12) ms. Pacing with a short or long atrioventricular interval gave similar maximum heart rates, oxygen uptake at the anaerobic threshold, end tidal pressure of carbon dioxide or oxygen pulse (a measure of stroke volume). Pacing with either a short or long atrioventricular interval produced a significantly higher oxygen consumption and anaerobic threshold and less lactate production than VVI pacing. During exercise a short atrioventricular interval does not provide a better cardiopulmonary performance than a long atrioventricular interval.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=1216619Documentos Relacionados
- Benefits of dual-chamber pacemakers.
- Optimal atrioventricular delay at rest and during exercise in patients with dual chamber pacemakers: a non-invasive assessment by continuous wave Doppler.
- Evaluation of six minute walking test in patients with single chamber rate responsive pacemakers.
- Disturbance of peripheral microvascular fluid permeability by the onset of atrioventricular asynchrony in patients with programmable pacemakers.
- Driving licences for patients with cardiac pacemakers.