Multicenter evaluation of the novel ABN Western blot (immunoblot) system in comparison with an enzyme-linked immunosorbent assay and a different Western blot.

AUTOR(ES)
RESUMO

A new, modular Western blot (immunoblot) system for human immunodeficiency virus (HIV) antibodies (ABN WesPage; Wellcome) was compared with enzyme immunoassays (Wellcome, Behringwerke, and Abbott) and with a U.S. Food and Drug Administration (FDA)-licensed Western blot (DuPont) in a multicenter study. A total of 649 serum samples from HIV patients at different stages of the disease, as well as from high-risk patients, from patients with conditions unrelated to AIDS, and from healthy blood donors, were used in the evaluation along with nine seroconversion panels. For evaluation of Western blot reactivity, both Centers for Disease Control (CDC) and FDA criteria were used. With the DuPont Western blot as the reference assay, the overall sensitivity and specificity of the ABN WesPage were 100 and 99.1%, respectively, when indeterminate results were not taken into account and when both tests were interpreted in accordance with CDC criteria. The DuPont Western blot detected significantly more antibodies to pol and gag gene products than the ABN WesPage. The ABN WesPage showed a higher positive rate of detection of viral envelope band gp160. When both Western blots were interpreted in accordance with CDC criteria, the ABN WesPage and the DuPont Western blot yielded 9.3 and 10.4% indeterminate results, respectively. When the DuPont Western blot was interpreted in accordance with the manufacturer's instructions (FDA criteria), 25.7% of the samples tested were regarded as indeterminate. The choice of interpretation criteria is of paramount importance for the evaluation of HIV Western blot patterns.

Documentos Relacionados