Microdilution Antibiotic Susceptibility Test: Examination of Certain Variables
AUTOR(ES)
Tilton, R. C.
RESUMO
A semiautomated microdilution susceptibility test is described. The effect of certain parameters such as inoculum size, growth media, incubation conditions, and inoculum dispensing systems was studied. Both medium type and inoculum size caused significant variations in the minimum inhibitory concentrations (MIC) of certain antibiotic-organism combinations. No effect on MIC was observed as a function of incubator type. Efforts to read a reproducible MIC value in less than 12 h failed. A commercially available wire pronged inoculator was determined to be inaccurate and unsafe. Disposable dropper pipettes proved to be economical, accurate, and precise. Although a standard method for microdilution antibiotic susceptibility testing is not proposed, data are presented which show that future attempts at standardized procedures are mandatory if inter- and intralaboratory reliability is desired.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=379880Documentos Relacionados
- Analysis of Certain Variables in the Agar Dilution Susceptibility Test
- KTK MOTOR TEST: REVIEW OF THE MAIN INFLUENCING VARIABLES
- Comparison of Broth Microdilution, E Test, and Agar Dilution Methods for Antibiotic Susceptibility Testing of Campylobacter jejuni and Campylobacter coli
- Comparison of BIOGRAM and commercial microdilution antimicrobial susceptibility test systems.