Latex agglutination inhibition card test for gentamicin assay: clinical evaluation and comparison with radioimmunoassay and bioassay.

AUTOR(ES)
RESUMO

Gentamicin levels were determined in 100 serum specimens by a new latex agglutination inhibition card test, a radioimmunoassay (RIA), and a bioassay. Correlation coefficients determined by linear regression analysis demonstrated that the levels obtained by the latex agglutination inhibition card test had a high degree of correlation with the RIA and could be performed much faster and more economically when processing small numbers of specimens. The bioassay had a slightly lower degree of correlation with both the RIA and the latex test and was adversely influenced by concurrently administered antibiotics which could not be eliminated by beta-lactamase. When measuring gentamicin concentrations above 2 micrograms/ml, the coefficient of variation was less than 14% for the latex agglutination assay compared with 15% for the bioassay and 12% for RIA. The latex agglutination inhibition card test is a rapid, accurate, specific, and reproducible method for monitoring gentamicin levels in patients and is particularly applicable for laboratories processing small numbers of specimens.