Improved procedure for determination of flucytosine in human blood plasma by high-pressure liquid chromatography.

AUTOR(ES)
RESUMO

Several high-pressure liquid chromatography procedures for the determination of flucytosine in serum or plasma have appeared. Some of these suffer from significant disadvantages, and none was applicable in our routine clinical therapeutic-drug-monitoring laboratory. A new high-pressure liquid chromatography assay for flucytosine was therefore developed. A 100-microliter sample of plasma was treated with an aqueous 5-iodocytosine internal-standard solution, and the mixture was deproteinized with trichloroacetic acid. A portion of the protein-free supernatant was diluted with 0.1 M ammonium phosphate, and an aliquot of the resulting solution was injected into the high-pressure liquid chromatography system. Chromatography was performed on a strong-cation-exchange column with a mobile phase containing aqueous ammonium phosphate, phosphoric acid, methanol, and acetonitrile. Detection was at 254 nm. The assay was shown to be linear in the 10 to 200-micrograms/ml drug-concentration range. Forty other drugs were tested for potential interference with the assay, and none was found. For routine use, a single-point working standard containing 75 micrograms of flucytosine per ml was used, giving intraassay coefficients of variation at 50 and 150 micrograms/ml of 1.8 and 2.3% respectively, whereas the day-to-day coefficient of variation at 50 micrograms/ml was 10.0%. Advantages of the procedure include the small sample size, the use of a convenient and reliable internal standard, speed, and simplicity. The assay is highly suitable for routine clinical drug-analysis laboratories.