Implantação de um programa de controle de qualidade para sistemas de planejamento de tratamento computadorizados de acordo com o TRS 430 / Implementation of a quality assurance program for computerized treatment planning systems according to TRS 430

AUTOR(ES)
FONTE

IBICT - Instituto Brasileiro de Informação em Ciência e Tecnologia

DATA DE PUBLICAÇÃO

27/04/2006

RESUMO

This work presents the guidelines and necessary tests tom implement a quality assurance program for Eclipse 7.3.10 from Varian at Hospital das Clinicas, Sao Paulo University School of Medicine - Brazil, in accordance with the new IAEA publication TRS 430. The recommended tests for the TRS 430 air mainly classified into acceptance tests, commissioning (dosimetric and non-dosimetric tests), and routine tests. The IAEA document\ s recommendations are being implemented at the hospital for two Varian linear accelerators - Clinac 600C e Clinac 2100C. The acceptance tests verified \ hardware\ , integration of network systems, data transfer and \ software\ parameters. The results obtained are in a good agreement with the manufacturer\ s specifications. Measurements of absolute dose in several set-ups were made for the commissioning dosimetric tests. These data were compared to the absolute doses determined by the TPS. The great majority of the tests showed 90% to 80% of the analyzed data in acceptance levels, with a good agreement between the experimental data and the data determined by the TPS. Only settings with asymmetric fields presented significant discords, showing the need for a more detailed inquiry for these settings. The non-dosimetric commissioning tests have also presented excellent results, with virtually all the system tools and general performance in compliance with TRS 430. The acceptance criteria have been applied for a comparison between the values of MUs generated by TPS and the calculated manually ones. The beams have been characterized for Eclipse with data transferred from CadPlan and with data from recommissioning of accelerators, so for these tests it was found a difference of at least 3% for the conformal field shape for the data originated in the beams of recommissioning and at least 4% for the data proceeded from CadPlan. The tolerance level established by TRS 430 for this setting was 3%.

ASSUNTO(S)

algorithms comissionamento controle de qualidade dosimetry experimental data iaea linear accelerators optimization phantoms photon beams planning quality assurance quality control radiation doses radiotherapy recommendations sistemas de planejamento de tratamento computadorizados

Documentos Relacionados