Imaging soft samples with the atomic force microscope: gelatin in water and propanol.
AUTOR(ES)
Radmacher, M
RESUMO
We have imaged mica coated with thin gelatin films in water, propanol, and mixtures of these two liquids by atomic force microscopy (AFM). The elastic modulus (Young's modulus) can be tuned from 20 kPa to more than 0.1 GPa depending on the ratio of propanol to water. The resolution is best in pure propanol, on the order of 20 nm, and becomes worse for the softer samples. The degradation in resolution can be understood by considering the elastic indentation of the gelatin caused by the AFM tip. This indentation becomes larger and thus the contact area becomes larger the softer the sample is. Therefore this study may be used to estimate the resolution to be expected with an AFM on other soft samples, such as cells. Nondestructive imaging was possible only by imaging at forces < 1 nN. This was difficult to achieve in contact mode because of drift in the zero load deflection of the cantilever, supposedly caused by temperature drift, but straightforward in tapping mode.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=1236243Documentos Relacionados
- Imaging biological structures with the cryo atomic force microscope.
- Atomic force microscope imaging contrast based on molecular recognition.
- Structural activity of a cloned potassium channel (ROMK1) monitored with the atomic force microscope: The “molecular-sandwich” technique
- Atomic force microscopy of DNA and bacteriophage in air, water and propanol: the role of adhesion forces.