Evaluation of COBAS AMPLICOR (Roche): Accuracy in Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by Coamplification of Endocervical Specimens
AUTOR(ES)
Livengood, Charles H.
FONTE
American Society for Microbiology
RESUMO
We evaluated further the accuracy of the COBAS AMPLICOR (Roche) (CA) PCR-based system in detection of Chlamydia trachomatis and Neisseria gonorrhoeae in endocervical specimens. Endocervical specimens collected for any indication for testing for C. trachomatis and N. gonorrhoeae among a university hospital health system population were included. Testing for C. trachomatis was done by two PCR methods, CA and manual microwell AMPLICOR (Roche) (MWA), and by culture; testing for N. gonorrhoeae was done by CA and culture. Discrepancy resolution was performed. Reproducibility testing and hands-on labor time measurements for CA were done. Among 654 C. trachomatis samples, the prevalence of true positivity was 9.2%, and among the 618 N. gonorrhoeae samples, the prevalence of true positivity was 4.4%. For detection of C. trachomatis, the sensitivity, specificity, and negative and positive predictive values were, respectively, as follows for each test: CA, 93.3, 99.7, 99.3, and 96.4%; MWA, 91.7, 99.7, 99.2, and 96.5%; and culture, 65.0, 100, 96.6, and 100%. For detection of N. gonorrhoeae those values were as follows: CA, 96.3, 100, 99.8, and 100%; and culture, 92.6, 100, 99.7, and 100%. Hands-on labor time for each clinical result was estimated to be at 7.5 min. The prevalence of inhibitory specimens was 3.5%, including two positive C. trachomatis samples which would have been missed otherwise. The direct cost of each clinical result with CA was estimated to be $9.09. Our methods include a diverse range of indications for testing among women, using endocervical swabbing samples, 2 M sucrose phosphate transport medium, and discrepancy resolution for comparison. Under our test conditions, the CA system is an accurate, rapid, and cost- and labor-efficient method for detection of C. trachomatis and N. gonorrhoeae.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=88262Documentos Relacionados
- High Frequency of Competitive Inhibition in the Roche Cobas AMPLICOR Multiplex PCR for Chlamydia trachomatis and Neisseria gonorrhoeae
- Validation of Roche COBAS Amplicor Assay for Detection of Chlamydia trachomatis in Rectal and Pharyngeal Specimens by an omp1 PCR Assay
- Detection of Neisseria gonorrhoeae and Chlamydia trachomatis in genitourinary specimens from men and women by a coamplification PCR assay.
- Evaluation of nucleic acid-based test (PACE 2C) for simultaneous detection of Chlamydia trachomatis and Neisseria gonorrhoeae in endocervical specimens.
- Multicenter Evaluation of the Fully Automated COBAS AMPLICOR PCR Test for Detection of Chlamydia trachomatis in Urogenital Specimens