Evaluation of an automated immunodiagnostic assay, VIDAS Rotavirus, for detection of rotavirus in fecal specimens.
AUTOR(ES)
Dennehy, P H
RESUMO
Four hundred sixty-four fecal specimens were tested for rotavirus by three immunoassays, VIDAS Rotavirus (bioMérieux Vitek, Hazelwood, Mo.), Rotaclone (Cambridge Biotech Corporation, Worcester, Mass.), and Pathfinder Rotavirus (Sansfi Diagnostics Pasteur, Chaska, Minn.). Twenty-seven discordant specimens were evaluated by electron microscopy. The sensitivities, specificities, positive predictive values, and negative predictive values were 98, 99.3, 98.7, and 99%, respectively, for VIDAS Rotavirus; 100, 99, 98.1, and 100%, respectively, for Rotaclone; and 100, 92.4, 87.3, and 100%, respectively for Pathfinder Rotavirus.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=263132Documentos Relacionados
- Evaluation of an automated immunodiagnostic assay system for direct detection of herpes simplex virus antigen in clinical specimens.
- Evaluation of the ImmunoCardSTAT! Rotavirus Assay for Detection of Group A Rotavirus in Fecal Specimens
- Comparison of nine commercial immunoassays for the detection of rotavirus in fecal specimens.
- Evaluation of an enzyme-linked immunosorbent assay for detection of Cryptosporidium spp. in stool specimens.
- Comparison of five enzyme immunoassays, electron microscopy, and latex agglutination for detection of rotavirus in fecal specimens.