Determinação de arsênio em águas contaminadas usando fluorescência de raios-X por energia dispersiva
AUTOR(ES)
Pataca, Luiz Carlos M., Bortoleto, Gisele G., Bueno, Maria Izabel M. S.
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2005-08
RESUMO
This work proposes a simple, fast and inexpensive method to determine As in natural waters, using X-ray fluorescence. 50 µL of each sample containing 100 mg L-1 of yttrium as internal standard were deposited over a 2.5 µm thickness MylarTM film. The samples were dried at 50 °C for 2 h. X-ray spectra were obtained using an EDXRF apparatus. The accuracy was determined by analyte addition/recovery and by comparison with Hydride Generation Atomic Absorption Spectrometry (HG AAS). A recovery of about 100% was obtained and the results were in good agreement with HG AAS. The method showed a relative standard deviation of 6.8% and a detection limit of 10.5 µg L-1 of As.
Documentos Relacionados
- Especiação e determinação de arsenio e selenio em amostras de aguas naturais por fluorescencia de raios X dispersiva em energia
- Composição mineral determinado por fluorescência de raios-X dispersiva de energia de frutos exóticos tropicais brasileiros
- Aplicação de alguns modelos quimiometricos a espectroscopia de fluorescencia de raios-X de energia dispersiva
- Aplicação de alguns modelos quimiométricos à espectroscopia de fluorescência de raios-X de energia dispersiva
- Microdensitometria de filmes de raios-X usando raios-X