Detection of Chlamydia trachomatis Endocervical Infections by Ligase Chain Reaction versus ACCESS Chlamydia Antigen Assay
AUTOR(ES)
Waites, K. B.
FONTE
American Society for Microbiology
RESUMO
Ligase chain reaction (LCR) was compared with ACCESS immunoassay for detection of chlamydial infections in females. Despite efforts to improve ACCESS performance by evaluation of specimens that were in the test performance “grey zone,” LCR remained more sensitive and was less expensive to perform. ACCESS had a sensitivity of 83.9%, a specificity of 99.7%, a positive predictive value of 96.3%, and a negative predictive value of 98.5%.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=85464Documentos Relacionados
- Diagnosis of Chlamydia trachomatis endocervical infections by a commercial polymerase chain reaction assay.
- Effect of Endocervical Specimen Adequacy on Ligase Chain Reaction Detection of Chlamydia trachomatis
- Detection of Chlamydia trachomatis in endocervical specimens by polymerase chain reaction.
- Clinical evaluation of a new polymerase chain reaction assay for detection of Chlamydia trachomatis in endocervical specimens.
- Detection of Chlamydia trachomatis in urine specimens from women by ligase chain reaction.