RESUMO
This work reports the validation of an analytical UV spectrophotometric method to assay acyclovir in hydrophilic matrices (assay and dissolution studies). The method was linear in the range between 2.5-20 µg mL-1, presenting a good correlation coefficient ( r = 0,9999). Precision and accuracy analysis showed low relative standart deviation (< 2.0 %) and a good recoveries percentual (98.9-100 %). The procedure was linear, accurate, and robust. The method is simple and cheap. It does not use polluting reagents and can be applied in dissolution studies, being an adequate alternative to assay acyclovir in hydrophilic matrices tablets.