Cord formation in BACTEC 7H12 medium for rapid, presumptive identification of Mycobacterium tuberculosis complex.
AUTOR(ES)
Yagupsky, P V
RESUMO
We evaluated cord formation in BACTEC 7H12 medium as a criterion for rapid identification of Mycobacterium tuberculosis complex. Kinyoun-stained smears, prepared from 270 radiometrically positive BACTEC 7H12 bottles, were examined independently by three observers. Smears from 93.2, 88.6, and 83.0% of the M. tuberculosis complex cultures were read as cord positive, and smears from 97.3, 97.8, and 99.5% of the mycobacteria other than M. tuberculosis cultures were read as cord negative by the three observers, respectively. There was 93.3% agreement between the observers. The presence of cords in BACTEC 7H12 medium can be a reliable criterion for rapid, presumptive identification of M. tuberculosis complex.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=267954Documentos Relacionados
- Cord Formation in BACTEC Medium Is a Reliable, Rapid Method for Presumptive Identification of Mycobacterium tuberculosis Complex
- Effect of egg yolk on growth of Mycobacterium tuberculosis in 7H12 liquid medium.
- Filtration of BACTEC 7H12B broth cultures for identification of Mycobacterium tuberculosis complex by AccuProbe.
- Cord Formation in MB/BacT Medium Is a Reliable Criterion for Presumptive Identification of Mycobacterium tuberculosis Complex in Laboratories with High Prevalence of M. tuberculosis
- Reliability of cord formation in BACTEC media for presumptive identification of mycobacteria.