Controlled evaluation of BacT/alert standard anaerobic and FAN anaerobic blood culture bottles for the detection of bacteremia and fungemia.

AUTOR(ES)
RESUMO

FAN medium was formulated to improve microbial recovery, particularly for fastidious microorganisms and for microorganisms causing sepsis in patients receiving antimicrobial therapy. In a controlled clinical evaluation performed at four university-affiliated hospitals, FAN anaerobic bottles were compared with standard anaerobic bottles for yield, speed of detection of microbial growth, and detection of septic episodes. A total of 10,431 blood culture sets were received; both anaerobic bottles of 7,694 blood culture sets were adequately filled with blood. Altogether, 925 isolates were recovered: 557 that were the cause of sepsis, 99 that were indeterminate as the cause of sepsis, and 269 contaminants. More Staphylococcus aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Escherichia coli (P < 0.02), and all microorganisms combined (P < 0.005) were recovered from FAN bottles; more nonfermentative gram-negative bacilli (P < 0.05), Torulopsis glabrata (P < 0.001), and other yeasts (P < 0.01) were recovered from standard bottles. Growth of S. aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Enterococcus faecalis (P < 0.025), streptococci other than Streptococcus pneumoniae (P < 0.01), and all microorganisms combined (P < 0.001) was detected earlier in standard bottles; growth of more isolates of E. coli (P < 0.05) and anaerobic bacteria (P < 0.01) was detected earlier in FAN bottles. The mean times to detection were 14.2 and 16.1 h for standard and FAN bottles, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)