Construção de uma cela de fluxo de longo caminho óptico para medidas espectrofotométricas
AUTOR(ES)
Melchert, Wanessa Roberto, Rocha, Fábio R. P., Raimundo Jr., Ivo M., Reis, Boaventura F.
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2008
RESUMO
A simple and low-cost flow cell with 30 cm optical path for spectrophotometric measurements is described. It presents desirable characteristics such as low attenuation of the radiation beam and internal volume (75 µL) comparable to that of a 1-cm conventional cell (80 µL). Despite the increase in optical path, the effect on sample dispersion was also similar to that attained in the commercial cell. The performance of the cell was assessed by the determination of phosphate based on the molybdenum blue method, yielding a linear response range between 0.05 and 0.8 mg L-1 phosphorus (r=0.999). The increase in sensitivity (30.4-fold) in comparison with that obtained with a conventional 1-cm flow cell agreed with that estimated by the Lambert-Beer law.
Documentos Relacionados
- Construção de uma cela de fluxo para medidas por espectrofotometria em fase sólida
- Determinação espectrofotométrica em fluxo de cloro em água usando célula de longo caminho óptico e multicomutação
- Construção e aplicação de uma cela espectrofotométrica de camada delgada para análises em fluxo
- Espectrofotometria de longo caminho óptico em espectrofotômetro de duplo-feixe convencional: uma alternativa simples para investigações de amostras com densidade óptica muito baixa
- Determinação espectofométrica de hipoclorito em alvejantes e cloro em águas de abastecimento empregando sistema em fluxo por multicomutação e células convencional e de longo caminho óptico