Computer-navigated Iliosacral Screw Insertion Reduces Malposition Rate and Radiation Exposure

AUTOR(ES)
FONTE

Springer-Verlag

RESUMO

Insertion of percutaneous iliosacral screws with fluoroscopic guidance is associated with a relatively high screw malposition rate and long radiation exposure. We asked whether radiation exposure was reduced and screw position improved in patients having percutaneous iliosacral screw insertion using computer-assisted navigation compared with patients having conventional fluoroscopic screw placement. We inserted 26 screws in 24 patients using the navigation system and 35 screws in 32 patients using the conventional fluoroscopic technique. Two subgroups were analyzed, one in which only one iliosacral screw was placed and another with additional use of an external fixator. We determined screw positions by computed tomography and compared operation time, radiation exposure, and screw position. We observed no difference in operative times. Radiation exposure was reduced for the patients and operating room personnel with computer assistance. The postoperative computed tomography scan showed better screw position and fewer malpositioned screws in the three-dimensional navigated groups. Computer navigation reduced malposition rate and radiation exposure.