Comparison of sensititre broth microdilution and agar dilution susceptibility testing techniques for meropenem to determine accuracy, reproducibility, and predictive values.

AUTOR(ES)
RESUMO

The stability, accuracy, reproducibility, and predictive value of Sensititre MIC panels containing meropenem (Merrem) were evaluated by using National Committee for Clinical Laboratory Standards (NCCLS)-recommended American Type Culture Collection (ATCC) strains and 110 selected strains of rapidly growing and fastidious aerobes and anaerobes with various degrees of susceptibility to meropenem. The NCCLS-recommended agar dilution method was used as a standard reference method. Meropenem-containing Sensititre MIC panels were monitored for their stabilities at room temperature and reproducibilities over 24 months by using six ATCC strains. Ninety-nine percent of the MICs of both meropenem and imipenem obtained for NCCLS-recommended ATCC strains were within the established ranges after 2 years. The overall agreement (+/- 1 twofold dilution) between the Sensititre and the agar dilution meropenem MICs was greater than 93%. The predictive value of meropenem MICs for indicating suspeptibility or resistance obtained by the Sensititre method was greater than 90%. No major or very major interpretive errors were observed, and only 5% of meropenem MICs were associated with minor interpretive errors. Problematic organisms were not observed. The Sensititre MIC panels containing meropenem offer a convenient and valid alternative to the NCCLS reference method for the susceptibility testing of potential pathogens likely to be recovered from mixed infections.