Comparison of four enzyme immunoassays for detection of human T-cell lymphotropic virus type 2 antibodies.

AUTOR(ES)
RESUMO

Four licensed enzyme immunoassay (EIA) kits for the measurement of antibody to human T-cell lymphotropic virus (HTLV) type 1, one from Organon Teknika Corp. (OTC), one from Cambridge Biotech Corp. (CBC), and two from Abbott Laboratories (the 1993 modification [Abb 93] and the 2.0 version licensed in 1995 [Abb 95]), were evaluated for sensitivity and specificity in the detection of HTLV type 2 antibody, and the results were compared with those previously obtained with earlier kit versions. The CBC, Abb 95, Abb 93, and OTC kits had sensitivities of 99.7, 97.6, 96.8, and 96.2%, respectively, compared with sensitivities of 89.1 and 60% for the Abbott and CBC (previously DuPont) kits, respectively, licensed in 1988. Thus, the abilities of commercial kits to detect HTLV antibody have improved. The relative specificities of the CBC, Abb 95, Abb 93, and OTC kits with negative blood donor specimens that had been reactive with the 1988 CBC EIA kit were 92.9, 64.5, 78.8, and 62.6%, respectively. Compared with those of the 1988 versions, the specificity of the Abbott EIA has decreased and the specificity of the CBC kit has been significantly improved.

Documentos Relacionados