Comparative study of test kits for measurement of rheumatoid factors by the latex fixation test.
AUTOR(ES)
Klein, F
RESUMO
The performance of 14 test kits for the measurement of rheumatoid factors by the latex fixation test was investigated; all results were expressed in IU/ml. Reproducibility and properties of control sera were also studied. It was found that half of the kits did not satisfactorily detect rheumatoid factor levels lower than 25 IU/ml. Most kits performed reasonably well in practice. Positive control sera only rarely met strict requirements. The relative costs did not always have a clear correlation with the quality of the results. The variation caused by differences between kits became acceptable when the results were expressed in IU/ml.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=1004238Documentos Relacionados
- Comparative Study of the Haemagglutination Test and a Simple Latex-Fixation Test for the Detection of the Rheumatoid Factor
- Measurement of amikacin in serum by a latex agglutination inhibition test.
- Measurement of antibodies to varicella-zoster virus by using a latex agglutination test.
- Evaluation of the gonococcal complement-fixation test.