Arc-to-arc mini-sling 1999: a critical analysis of concept and technology
AUTOR(ES)
Palma, Paulo
FONTE
International braz j urol
DATA DE PUBLICAÇÃO
2011-04
RESUMO
PURPOSE: The aim of this study was to critically review the Arc-to-Arc mini-sling (Palma's technique) a less invasive mid-urethral sling using bovine pericardium as the sling material. MATERIALS AND METHODS: The Arc-to-Arc mini-sling, using bovine pericardium, was the first published report of a mini-sling, in 1999. The technique was identical to the "tension-free tape" operation, midline incision and dissection of the urethra. The ATFP (white line) was identified by blunt dissection, and the mini-sling was sutured to the tendinous arc on both sides with 2 polypropylene 00 sutures. RESULTS: The initial results were encouraging, with 9/10 patients cured at the 6 weeks post-operative visit. However, infection and extrusion of the mini-sling resulted in sling extrusion and removal, with 5 patients remaining cured at 12 months. CONCLUSION: The Arc-to-Arc mini-sling was a good concept, but failed because of the poor technology available at that time. Further research using new materials and better technology has led to new and safer alternatives for the management of stress urinary incontinence.
Documentos Relacionados
- A critical analysis of mini peer assessment tool (mini-PAT)
- Invasive Pneumococcal Infections in Denmark from 1995 to 1999: Epidemiology, Serotypes, and Resistance
- Science and Technology Studies and the Historiography of Psychology: Towards a Critical Analysis
- Mini-invasive collagen sling in the treatment of urinary incontinence due to sphincteric incompetence
- The Mersilene mesh sling--a new concept in ptosis surgery.