Determinação de ácido valpróico em soro por cromatografia líquida de alta eficiência com detector de arranjo de diodos (CLAE-DAD), após derivatização com brometo de fenacila

AUTOR(ES)
FONTE

Química Nova

DATA DE PUBLICAÇÃO

2009

RESUMO

Valproic acid (VA) is a drug used to control seizures in several epileptic conditions. In VA pharmacotherapy, therapeutic drug monitoring is recommended to obtain adequate seizure control and avoid toxicity. The aim of this study was to validate a method for the determination of valproic acid in serum, employing high performance liquid chromatography with diode array detection (HPLC-DAD), after derivatization with phenacyl bromide. The calibration curve (y=0.0133x-0.0025) presented good linearity with r²=0.9999. Accuracy (101-115%), intra-assay precision (4.53-8.15%) and inter-assay precision (3.15-6.77%) were acceptable. The quantification limit was 2.0 µg/mL. The method presented similar results to enzyme immunoassay.